Abstract

We have measured the hysteresis in output voltage arising from loss of power for three Fluke 732B dc voltage references. Continuously powering the internal temperature-controlled oven provides the best accuracy with such Zener-diode-based dc voltage references. Batteries are used to provide power continuity when shipping between laboratories. We have shown that for these particular instruments, an apparent hysteresis induced by humidity is likely to dominate any hysteresis arising from loss of battery power when transit between laboratories takes three days or more. Our conclusion is that as much, if not more, care needs to be taken when dealing with humidity than with loss of battery power when shipping between laboratories. When the highest accuracy of dc voltage comparison is required, it becomes necessary to apply voltage corrections using the measured values of humidity coefficients and the relative humidity occurring during transit.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.