Abstract

The heating of tissues around implants during MRI can pose severe health risks, and careful evaluation is required for leads to be labeled as MR conditionally safe. A recent interlaboratory comparison study has shown that different groups can produce widely varying results (sometimes with more than a factor of 5 difference) when performing measurements according to current guidelines. To determine the related difficulties and to derive optimized procedures, two different generic lead structures have been investigated in this study by using state-of-the-art temperature and dosimetric probes, as well as simulations for which detailed uncertainty budgets have been determined. The agreement between simulations and measurements is well within the combined uncertainty. The study revealed that the uncertainty can be kept below 17% if appropriate instrumentation and procedures are applied. Optimized experimental assessment techniques can be derived from the findings presented herein.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call