Materials and methods Three cadaveric vertebrae T7, L1, and L3 were scanned with a 3D medical ultrasound system. The rotation angles of the cadaveric vertebra were recorded during the experiment. Nine sets of ultrasound data, from 0 to 40° with 5° increments, were recorded from each vertebra (27 sets in total). An in-house program was used to reconstruct the 3D vertebra images. The rotation of each reconstructed vertebra was determined by the angle between the line going through either the centres of laminae (L-L) or the centres of transverse processes (TPTP) and a reference vertical plane. This reference plane was defned from the position sensor parallel to the surface of the transducer. Three raters, who were blinded with the rotation information, used the images to measure the rotation in 3 sessions. In each session, the raters used the mouse pointer to select L-L or TP-TP according to their knowledge of vertebral anatomy. The program received the 3D coordinates of these points and calculated the VR. Intra-class correlation coefficients (ICCs) (two-way random and absolute agreement) were used to calculate the intraand inter-reliability. The mean absolute difference (MAD±SD) and the range of difference (RD) between the true values and the average measurements of each rater were also computed to evaluate the accuracy of methods. Results When rotation was greater than 30° for both L1 and L3, all raters found it difficult to determine one of the lamina areas because it could not be displayed due to ultrasound blocking. Therefore, the corresponding measurements were excluded. The intra-reliability (L-L, TP-TP) for the three raters were (0.987, 0.991), (0.989, 0.998) and (0.997, 1.000), respectively; meanwhile, the inter-reliability were 0.991 for (L-L) and 0.992 for (TP-TP). All ICC values were greater than 0.98 indicating both methods were highly reliable. The MAD±SD values (L-L, TP-TP) for the three raters were (1.5±0.3°, 1.2±0.2°), (1.6±0.3°,1.3 ±0.3°), and (1.7±0.5°, 0.9±0.2°), respectively. The RD (L-L, TP-TP) were (0-4.5°, 0-3.5°), (0-5.1°, 0-4.3°), and (0-5.1°, 0-2.8°) for the three raters, respectively.

Highlights

  • Vertebral rotation (VR) is an important parameter to evaluate the severity of scoliosis and predict the progression

  • This study investigated a new three-dimensional (3D) nonionizing method to measure VR based on either laminae or transverse processes

  • Published: 19 January 2015 doi:10.1186/1748-7161-10-S1-O40 Cite this article as: Vo et al.: Measurement of vertebral rotation using a three-dimensional ultrasound image

Read more

Summary

Objective

Vertebral rotation (VR) is an important parameter to evaluate the severity of scoliosis and predict the progression. This study investigated a new three-dimensional (3D) nonionizing method to measure VR based on either laminae or transverse processes

Materials and methods
Results
Conclusions
Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.