Abstract

A new reverse-phase high-performance liquid chromatographic (HPLC) procedure has been developed for the quantitation of vancomycin and its crystalline degradation product, CDP-1. The new HPLC procedure involves a liquid-liquid extraction pretreatment procedure and an HPLC internal standard, vitamin B12. The new HPLC procedure was used to measure 50 renally impaired patient samples. Samples were than analyzed by the monoclonal antibody Emit vancomycin assay, and the polyclonal antibody fluorescence polarization immunoassay (FPIA) vancomycin assay, and the results were compared. The Emit vancomycin assay correlated well with HPLC, but FPIA quantitated higher than Emit and HPLC. We conclude that the polyclonal antibody FPIA assay detects CDP-1 in the samples of renally impaired patients, accounting for an average of 10% of the FPIA/Emit bias, and should therefore be used with caution when monitoring vancomycin levels in renally impaired patients.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.