Abstract

Aqueous ethanol wet-bath simulator solutions are used to perform calibration adjustments, calibration checks, proficiency testing, and inspection of breath alcohol instruments. The Toxicology Bureau of the New Mexico Department of Health has conducted a study to estimate a measurement of uncertainty for the preparation and testing of these wet-bath simulator solutions. The measurand is identified as the mass concentration of ethanol (g/100 mL) determined through dual capillary column headspace gas chromatography with flame ionization detector analysis. Three groups were used in the estimation of the aqueous ethanol wet-bath simulator solutions uncertainty: GC calibration adjustment, GC analytical, and certified reference material. The standard uncertainties for these uncertainty sources were combined using the method of root-sum-squares to give uc = 0.8598%. The combined standard uncertainty was expanded to U = 1.7% to reflect a confidence level of 95% using a coverage factor of 2. This estimation applies to all aqueous ethanol wet-bath simulator solution concentrations produced by this laboratory.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.