Abstract

The detection efficiencies of tritium samples by using liquid scintillation counter with hydrophilic plastic scintillator (PS) was approximately 48% when the sample of 20μL was held between 2 PS sheets treated by plasma. The activity and count rates showed a good relationship between 400Bq to 410KBqmL−1. The calculated detection limit of 2min measurement by the PS was 13BqmL−1 when a confidence was 95%. The plasma method for PS produces no radioactive waste.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.