Abstract

A detergent-scintillation liquid mixture applied to measure low specific activity of tritium in natural water was studied. The DYS-1 low level liquid scintillation counter designed and manufactured by our institute was employed. In comparing the Triton X-100 scintillation liquid mixture with the dioxane-based-scintillation liquid, a better formula for Triton X-100 scintillation liquid mixture was determined, the mixture possesses the quality of high water content; high efficiency and low back-ground in measuring tritium in water. Chemiluminescence of the Triton X-100 scintillation liquid mixture can be totally de-excited in short time. It can be employed at ambient temperature 11–28°C. For 20ml sample in quartz vials, counting efficiency is 15% with a background 2.17 cpm, Y=31 TU (t=30 min).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.