Abstract
The differential arrangement of two pneumotachographs in a gas blow-by system, upstream and downstream from the patient connection port, permits accurate measurement of the tracheal air flow without introduction of flowmeter dead-space. The method does not require but will accept a constant background gas flow rate and permits humidification and easy switching of the inhalant gases. The admissible background flow rates limiting measurement errors below 5% in the range of tidal volumes relevant for newborns for two sizes of commercially available pneumotachographs were found to be -3 L/min and 0-12 L/min, respectively. The lack of restrictions imposed on the properties of the background flow source makes the differential system especially suitable for measurement of ventilation and pulmonary mechanics in spontaneously breathing or mechanically ventilated newborn infants.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.