Tin is reduced to stannane by sodium tetrahydroborate solution, passed through a heated silica tube and measured by atomic absorption specirometry at 286.33 nm. The detection limit is 0.9 ng ml -1 and the r.s.d. is 4.1% at 0.01 μg ml -1. Interferences are discussed and the sample decomposition procedure evaluated. Typical results obtained for macroalgae. fish and molluscs are given.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.