Abstract

The relative yield of 3γ to 2γ annihilation was measured for a positron emitter, 22Na, with the new scintillator detectors lanthanum chloride (LaCl3:Ce) and lanthanum bromide (LaBr3:Ce), which had been characterised for comparison with high-purity germanium (HPGe) and sodium iodide (NaI(Tl)) detectors. The information obtained from the ortho-positronium 3γ decay in positron emission tomography (PET) can be a measure of the oxygen content in biological tissues by determination of this relative yield. However, it requires high resolution spectroscopy and detection efficiency. Characterisation of the new generation of scintillator detectors determines whether they could replace conventional scintillators and semiconductors. A series of experiments was carried out with different samples in order to study the effect of ortho-positronium formation. The peak-to-peak and the peak-to-valley methods were compared in the measurement of the relative yield of 3γ to 2γ annihilation.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.