Abstract

A Talbot interferometer with circular grids is used to measure the temperature profile inside a premixed laminar flow slot burner used in an atomic absorption spectrophotometer. The experimentally measured results using the Talbot interferometry are compared with the values measured by a platinum-platinum-rhodium thermocouple with accuracy of measurement of ±0.2% full scale ±four digits and good agreement is found. The effect on the temperature profile of aspirating water through the nebulizer is also investigated.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.