Abstract

To determine noise intensity during middle-ear aspiration in order to evaluate whether levels can be potentially harmful. In this prospective, observational study, middle-ear effusion was aspirated following myringotomy using a suction instrument with a probe tube microphone. Sound pressure levels and duration were measured, and frequency domain analysis was performed. Forty-four ears were analysed, consisting of 20 with mucoid effusion, 11 with serous effusion and 13 with no effusion. Maximum peak sound intensity ranged from 84 to 157dB. Half of the ears (50 per cent) were exposed to greater than 140dB; of these, 82 per cent were exposed for longer than 0.2ms (range, 0.05-14ms). There was no significant difference in sound pressure level between ears with mucoid and serous effusion; however, ears with mucoid effusion required longer suction times (p<0.0030). In addition, peak intensity was greater for ears with mucoid effusion versus those with serous or no effusion (p<0.0001). Middle-ear aspiration during myringotomy caused noise levels within a potentially harmful range.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.