Abstract

Diagnosis of subluxation and dislocation of the distal radioulnar joint (DRUJ) is difficult. In order to measure the shift between radius and ulna, we evaluate a method with a custom platform in this study. Cadaver specimen was fixed on a custom platform, and the volar and dorsal translation of the ulnar head was measured under different stress forces. We built a subluxation model via a surgical procedure that removed components of the radioulnar joint, and the translation at varying stress levels was measured again. We find at the same stress level, the normal model results in significantly less translation than the subluxation model at the majority of the stress forces, and particularly with volar forces. Our results indicate that this measurement is a potential method to evaluate the stability of the DRUJ.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.