Abstract
Nowadays, ionizing radiation beams with relatively large transverse dimensions are widely used not only for research purposes but also in technological processes related to metal and food treatment, as well as medical instrument sterilization. Thus, control of beam characteristics is required, and in particular, the intensity distribution in the transverse plane must be accurately measured. The multiangle scanning method, based on mathematical transformations, allows overcoming the limitations of modern detectors used for these purposes. In this study, the method was applied to an electron beam. A detection system based on scintillating fiber was used. The beam profile was measured using the proposed method and a conventional detector — a film dosimeter. Experimental results allowto conclude that the beam intensity distribution obtained using the multiangle scanning method is close to the results obtained using a film dosimeter, which is most often used in medical institutions. Thus, we can consider the multiangle scanning method as an effective approach for determining the intensity distribution of an electron beam.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.