Abstract

A noninvasive method using an inversion recovery turbo-FLASH for dynamic measurement of the arterial input function represented by the bolus passage of Gd-DTPA in the descending aorta is presented, and the results are compared with the input function obtained by arterial blood samples. A good accordance between the two input functions was found, indicating that it is possible to measure the input function to the myocardium using MRI. A variation between the two concentration curves of 5% at upslope, 2.7% at peak point, and < 7% at downslope was found. The study also indicates that a short inversion time < 250 ms has to be used to ensure correct measurement of peak concentration.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.