Abstract

A new determination of the 109Cd half-life was made by a time series of measurements of an aqueous sample using a re-entrant type ionisation chamber. The measurement campaign covered a period of 6 years or approximately 4.7 half-lives of 109Cd. The resulting value of 462.1 (3) days is in good agreement with the recently published values of 462.29 (30) days and 462.3 (8) days. This new half-life determination will allow evaluators to specify a recommended value of the 109Cd half-life making it more accurate and precise.