Abstract

An electrospray mass spectrometric method for the quantification of the depolarizing neuromuscular blocking agent succinylcholine (SUX) is described. An extraction method compatible with direct infusion inlet was developed and leads to an analysis cycle time of 7–8 min instead of 25 min that would be required for HPLC inlet. SUX was extracted from human plasma on C1 solid-phase cartridges and was analyzed using positive ion electrospray tandem mass spectrometry (ESI-MS/MS). SUX plasma concentrations were determined by a stable isotope dilution assay using hexadeuterosuccinylcholine diiodide (SUXd6) as the internal standard. The calibration curve was prepared using the ratio of intensities of the major product ions in the collision-induced dissociation spectrum for known concentration ratios of SUX and SUXd6 in plasma. Calibration curves for the quantification were linear from 25 to 4000 ng/ml. For intraday precision, CV were ≤6% and accuracy ranged from 98 to 103%. For the interday precision, CV were ≤10% and accuracy ranged from 90 to 102%. This method is specific, sensitive, reproducible, and practical in a clinical setting.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.