Abstract

The thin foil calorimeter technique described earlier is extended to detail specific heats of polymers near transition temperatures. A sample film is rolled up with a metal foil heating element into a cylindrical configuration. Starting with the roll at thermal equilibrium, a constant current is passed through the foil. During a finite period after heating begins, the interior of the roll heats adiabatically. The temperature rise rate, sensed by a thin foil thermocouple, is monitored by a microvolt amplifier-recorder system. A typical 35 sec run shows a linear temperature rise of about 7×10−3 K, with a noise level of the order of 15×10−6 K. Resultant specific heat values are estimated to be precise to within 1%. Resistance thermometry gives corresponding mean sample temperatures with a precision of the order of 0.01 K. Although specific heat values obtained for polytetrafluoroethylene mostly agree with published adiabatic calorimeter data, remarkably different results are obtained in the transition temperature region extending from 280 to 310 K. Where adiabatic calorimeter data show a large enthalpy peak extending over a 15 K temperature interval, foil calorimeter data show a relatively minor but sharp discontinuity near 293 K.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.