Abstract

The performance test is an important and necessary work for the micro-CT (computed tomography) system. The focal spot size of the micro focus X-ray tube is measured. The method of measuring the spatial resolution of micro-CT is introduced. A line-pair resolution of 28.2 lp/mm at the 10% modulation transfer function (MTF) level can be achieved with 14.7 μm spot size, 12.3 μm voxel size and a 25 mm field of view. In addition, a tungsten wire with the diameter of 5 μm can be detected by the system.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.