Abstract

We have used laser-produced plasmas as a broadband source of soft x rays to measure the normal incidence reflectance of multilayer mirrors in the 10–25 nm spectral region. The measurements have a spectral resolution of 0.03 nm and a reflectance resolution of 10%. Measurements made on a Mo/Si multilayer show excellent agreement with results obtained using a synchrotron and indicate a normal incidence peak reflectance of over 50% at 15 nm. By repeating the reflectance measurement at different positions across a single 7.5 cm mirror we determined multilayer uniformity as a function of position and we relate this dependence to the geometry of the deposition process.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.