Abstract

In external beam radiation therapy (EBRT), skin dose measurement is important to evaluate dose coverage of superficial target volumes. Treatment planning systems (TPSs) are often inaccurate in this region of the patient, so in vivo measurements are necessary for skin surface dose estimation. In this work, superficial dose distributions were measured using radiochromic translucent poly(vinyl alcohol) cryogels. The cryogels simultaneously served as bolus material, providing the necessary buildup to achieve the desired superficial dose. The relationship between dose to the skin surface and dose measured with the bolus was established using a series of oblique irradiations with gantry angles ranging from 0° to 90°. EBT‐2 Gafchromic film was placed under the bolus, and the ratio of bolus‐film dose was determined ranging from 0.749 ± 0.005 to 0.930 ± 0.002 for 0° and 90° gantry angles, respectively. The average ratio over 0–67.5° (0.800 ± 0.064) was used as the single correction factor to convert dose in bolus to dose to the skin surface. The correction factor was applied to bolus measurements of skin dose from head and neck intensity‐modulated radiation therapy (IMRT) treatments delivered to a RANDO phantom. The resulting dose distributions were compared to film measurements using gamma analysis with a 3%/3 mm tolerance and a 10% threshold. The minimum gamma pass rate was 95.2% suggesting that the radiochromic bolus may provide an accurate estimation of skin surface dose using a simple correction factor. This study demonstrates the suitability of radiochromic cryogels for superficial dose measurements in megavoltage photon beams.

Highlights

  • Skin dosimetry is an important aspect in radiotherapy of superficial targets considering that Treatment planning systems (TPSs) are inaccurate in this region of the patient

  • A comparison of EBT-2 Gafchromic film and ferrous benzoic xylenol orange (FBX)-poly(vinyl alcohol) cryogels (PVA-C) radiochromic bolus suggests that the latter may provide an accurate estimation of skin surface dose distribution using a simple correction factor

  • Radiochromic bolus may be used in place of more traditional forms of bolus to perform in vivo dosimetry in regions where the skin dose is important

Read more

Summary

Objectives

The purpose of this work was to evaluate the ability of FBX-PVA-C radiosensitive bolus material to measure skin dose during radiotherapy

Methods
Results
Conclusion