Abstract

Abstract In certain situations, such as in viscous liquids or nano-porous media, the transversal relaxation time T 2 may become so short that the pulse duration or the detector dead time is no longer negligible. These cases raise fundamental questions about relaxation during the radio-frequency pulses used in classical CPMG or Inversion-Recovery sequences. For an ideal system having a single relaxation time, we examine the effect of the finite pulse duration on the magnetization decay after the pulses occurring in FID, CPMG and Inversion-Recovery sequences. We solved analytically the Bloch equations during and after the pulses and compared the theoretical predictions with experimental data. Finally, we propose approximate simple expressions to correct the sequences for the magnetization attenuation during the pulses. IR curves are affected by transversal relaxation during the pulses, yielding asymmetric curves even if T 1 is very large. The magnetization decay obtained during a CPMG sequence is not affected by relaxation during the pulses. This is valid provided the time origin is chosen in the middle of the first pulse.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call