Abstract
For the measurement of ultrasonic shock waves with amplitudes near 40 MPa, two different hydrophones were constructed. The first is a high-bandwidth PVDF membrane hydrophone with a capacitively coupled signal. The second is a small acousto-optic fiber hydrophone which measures the shock-wave-induced variation of the refractive index of a liquid at the front end of the fiber. The properties of the hydrophones are compared. The first, with a sensitivity of 6 mV/MPa, demonstrates high durability and a 20-MHz bandwidth. The second has equally good characteristics and additionally measures the creation and the time dependence of cavitation bubbles in the liquid. >
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.