We report on a rubidium-xenon (Rb-Xe) polarization unit for the continuous production of large quantities of hyperpolarized (129)Xe. The unit includes two diagnostic systems which enable the absolute measurement of both the (85)Rb and (129)Xe polarization in situ and at high temperatures. The Rb diagnostic system allows the measurement of one- or two-dimensional images of the absolute Rb polarization and thus enables the experimental study of light penetration into the optical pumping cell. The equilibrium Xe polarization measured in the optical pumping cell and in the freezing unit is typically approximately 20%, under optimal flowing conditions, and this is much lower than the volume-averaged rubidium polarization.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.