Abstract

The measurements of the K X-ray intensity ratio I(Kα2/Kα1), I(Kβ1/Kα1) and I(Kβ/Kα) for elements V, Mn, Zn, Tc, Ru, Cd, Xe, Ba, Cs, Hg and Rn were experimentally determined both by photon excitation, in which 59.5keV γ-rays from a 241Am and 123.6keV γ-rays from a 60Co were used, and following the radioactive decay of 51Cr, 55Fe, 67Ga, 99Tc, 111In, 131I, 133Ba, 133Xe, 137Cs, 201Tl and 226Ra. K X-rays emitted by samples were counted by a Si(Li) detector with resolution 160eV at 5.9keV. Obtained values were compared with the theoretical values. It was observed that present values agree with the previous theoretical and other experimental results.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.