Abstract

AbstractWithin the last 2 years our consortium has developed, built and tested an atomic force microscope for planetary science applications, in particular for the study of Martian dust and soil. An array of eight cantilevers and tips provides redundancy in case of tip or cantilever failure. Images can be recorded in both static and dynamic operation modes. As we plan to investigate Martian dust, our interest focuses on the behaviour of the instrument when measuring loose particles in the above‐mentioned modes. During scanning, tip contamination with a particle occurs quite frequently, altering the quality of the images. Before changing the cantilever, reverse‐imaging the contaminated tip on a tip calibration sample will be performed in order to increase the scientific throughput. We present the results of our test measurements on respirable α‐quartz. Copyright © 2002 John Wiley & Sons, Ltd.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.