Abstract

This paper describes a method to measure the product of Prepl Pwall correction factors for ionization chambers and presents our measured values of Prepl Pwall for Markus plane-parallel chambers in electron beams. It is shown that the measured values of Prepl Pwall can be fitted to an equation, Prepl Pwall = c1 + c2 R50 + c3 (R50)2, for Markus chambers at the new reference depth for electron beams (6 MeV < or = nominal energy E < or = 20 MeV). We also present our measured values of Prepl Pwall for NACP and Markus chambers in a water phantom irradiated in a 60Co beam.