The indirect measurement of pleural pressure in neonates is obtained from measurements of esophageal pressure (Pes) with either a liquid-filled catheter or an esophageal balloon-catheter system. The purpose of this investigation was to assess the validity of the water-filled esophageal catheter by comparing the simultaneous changes in Pes and airway opening pressure (Pao) during occluded respiratory efforts. Equal changes in Pes and Pao under this condition indicate that Pes is a valid measurement of pleural pressure. In six healthy unsedated term neonates (aged 2-3 days) we measured Pes in the lower third of the esophagus with a water-filled catheter of eight French gage (FC), which has a 2-mm internal diameter. During occlusions, changes in Pes and Pao were almost identical in magnitude and timing in each body position studied (right lateral, prone, and supine). We conclude that the water-filled 8-FG esophageal catheter gives an accurate measurement of pleural pressure changes in healthy neonates.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.