Abstract
We investigate novel methods for the quantification of platelet responsiveness that are suited for implementation in lab-on-a-chip devices. Magnetic beads are convenient carriers for rapid capture and manipulation of biological cells in a miniaturized system. In this article, we demonstrate that antibody-coated magnetic beads can be used to quantify platelet responsiveness. We use anti-CD62P coated beads to capture activated platelets from samples stimulated with a PAR-1 specific agonist SFLLRN, also known as thrombin receptor activator peptide. The responsiveness of the platelets is analyzed via the remaining unbound platelets in the solution and compared to a reference method in which the number of activated platelets is analyzed via fluorescent labeling. The effective concentrations for platelet activation are in agreement for the two assay types, proving that platelet responsiveness can be quantified using antibody-coated magnetic beads. We discuss the outlook for application in lab-on-a-chip devices.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.