Abstract

Plasma viscosity (PV) is used in the investigation of hyperviscosity syndrome. We have evaluated the performance of a new bench-top free oscillation rheometer, ReoRox from Medirox, and established a new reference range. Performance of the free oscillation rheometry (FOR) method was evaluated examining within-run and between-run imprecision, duration of measurement and operator simplicity. Furthermore, influence of storage on the PV measurements was assessed and a reference range based on plasma from 585 healthy men and women was established. Imprecision of the free oscillation rheometer was comparable to other viscometers and performance of the instrument was satisfying. The non-parametrical reference range established was 1.22-1.44 mPa s at 37℃ (2.5th and 97.5th percentiles). We have established a FOR PV reference range to support clinical decision-making. Furthermore, we found high pre-analytical stability of the samples and an analytical imprecision comparable to other PV methods.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.