Abstract
The CRESST experiment, aiming at the direct detection of WIMPs via nuclear recoils, is currently using scintillating CaWO4 crystals. The WIMP–nucleus cross section for elastic scattering as well as the scintillation efficiency differ considerably for recoils from Ca, W and O in these crystals. Therefore a discriminating detector calibration is essential in order to improve WIMP parameter claims. At the tandem accelerator of the Maier-Leibnitz-Laboratory (MLL) in Garching, Germany, a neutron scattering facility is operated for the determination of the individual quenching factors (QF) in the bulk of a CaWO4 crystal to better understand the detector response to neutron background and a possible WIMP signal. First measurements at room temperature reveal QF(O)=7.8±0.3% (recoil energy 1.0–2.2MeV), QF(Ca)=6.3±1.6% (recoil energy 0.4–1MeV), QF(W)<3.0% (2σ, recoil energy 0.1MeV).
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.