Abstract

We report a novel system to analyze atmospheric nitrophenols (NPs). Rain or air sample extracts (1 mL) are preconcentrated on a narrow bore (2 mm) aliphatic anion exchanger. In the absence of strong retention of NPs exhibited by aromatic ion exchangers, retained NPs are eluted as a plug by injection of 100 microL of 0.1 M Na(2)SO(4) on to a short (2 x 50 mm) reverse phase C-18 column packed with 2.2 mum particles. The salt plug passes through the C-18 column unretained while the NPs are separated by an ammonium acetate buffered methanol-water eluent, compatible with mass spectrometry (MS). The eluted NPs are measured with a long path Teflon AF-based liquid core waveguide (0.15 x 1420 mm) illuminated by a 403 nm light emitting diode and detected by a monolithic photodiode-operational amplifier. The waveguide is rendered chemically active by suspending it over concentrated ammonia that permeates into the lumen. The NPs ionize to the yellow anion form (lambda(max) approximately 400 nm). The separation of 4-nitrophenol, 2,4-dinitrophenol, 2-methyl-4-nitrophenol, 3-methyl-4-nitrophenol, and 2-nitrophenol (these are the dominant NPs, typically in that order, in both rain and air of Houston and Arlington, TX, confirmed by tandem MS) takes just over 5 min with respective S/N = 3 limits of detection (LODs) of 60, 12, 30, 67, and 23 pg/mL compared to MS/MS LODs of 20, 49, 11, 20, and 210 pg/mL. Illustrative air and rain data are presented.