Abstract

Exposure dose rate to the tumor and surrounding cells during neutron beam irradiation in Boron Neutron Capture Therapy (BNCT) comes not only from heavy charged particles produced from the 10B(n,α)7Li nuclear reaction, but also from neutron-induced reactions with other biological elements in living tissue, as well as from gamma rays leaked from the reactor core. At Dalat Research Reactor, Vietnam, the neutron and gamma dose rate distribution inside a water phantom were measured by using activation method and Thermoluminescent Dosimeter (TLD) detector, respectively. The results showed that effective thermal neutron dose rate along the center line of the water phantom had a maximum value of 479 mSv h-1 at 1 cm in phantom and then decreases rapidly to 4.87 mSv h-1 at 10 cm. The gamma dose rate along the center line of the water phantom also reach its maximum of 4.31 mSv h-1 at 1 cm depth and decreases to 1.16 mSv h-1 at 10 cm position. The maximum biological tumor dose rate was 1.74 Gy-eq h-1, not high enough to satisfy the treatment requirement of brain tumors. However, the results of this work are important in supporting of BNCT study in the upcoming stages at Dalat Research Reactor.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.