Abstract The success of minimally invasive procedures under MR-guidance can be increased by the knowledge of the current needle pose. We hypothesize that a one-toone mapping exists between the needle orientation with respect to the static magnetic field and the cross-sectional shape of the needle’s susceptibility artifact. For this purpose, we derived a mathematical model, which describes the cross sectional geometry of the needle artifact. It is approximated by two ellipses. Certain parameters of these ellipses can be utilized for mapping the geometry of the needle artifact onto the needle orientation. The relation between the two ellipse parameters α (inclination of the semi-major axis) and b (length of the semi-minor axis) and the needle’s azimuth angle can be approximated by linear regression in a certain angle interval. A combination of these two ellipse parameters is suitable for estimating the needle’s azimuth angle within a range between 0° and 60°.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.