Abstract

This study describes measurements on secondary particles produced by a 290 MeV/n Spread Out Bragg Peak (SOBP) carbon beam. Microdosimetric distributions of secondary fragments from the SOBP carbon beam have been measured by using a new tissue equivalent proportional counter (TEPC) system at the Heavy Ion Medical Accelerator in Chiba of the National Institute of Radiological Sciences. The new TEPC system consists of a TEPC, two solid-state detectors (SSD) and a scintillation counter (FSC: forward scintillation counter). The SSDs and FSC can separately identify charged fragments and secondary neutrons produced by the incident carbon ions. Microdosimetric distributions were measured for secondary particles including neutrons produced by a body-simulated phantom consisting of various PMMA plates (thickness: 0, 34.81, 55.2, 60.95, 64.83, 95.03, 114.79, 124.69, 135.2 and 144.98 mm, respectively) to cover the SOBP (at 60-125 mm depth). The new system can separately determine produced fragments from the incident SOBP carbon beam in a body-simulated phantom.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.