We describe here the design of the modernized ion-optical system (IOS) of the standard residual gas analyzer (RGA). IOS electrodes are made of stainless steel. The electrodes are powered outside the vacuum chamber. Reducing the diameter of the entrance aperture to 0.6 mm made it possible to expand the measurement range. These design features of this system makes it possible to measure the mass-to-charge composition of beam plasma ions generated by electron beam evaporation of ceramics and refractory metals in the forevacuum pressure range under a prolonged exposure to high temperature. We present the results of the measured plasma composition during evaporation of aluminum oxide and zirconium ceramics, as well as of titanium.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.