Abstract

In order to evaluate the reliability of the ID ICP-MS method for the measurement of magnesium, zinc, and copper in human serum, we investigated the traceability, precision, trueness, and uncertainty of the method. This method traces the contents of magnesium, zinc, and copper in human serum to the standard materials NIST SRM3131a, SRM3168a, and SRM3114 respectively, thus completing the traceability to SI unit. The repeatability of this method for measuring magnesium, zinc, and copper in the human serum reference material GBW09152 was found to be 0.2%, 0.7%, and 0.6% (n = 9), respectively. The measurement, when employed to measure the magnesium, zinc, and copper in standard materials, had caused a maximum deviation of less than 0.88%, 1.35%, and 1.15%, respectively. The measurement results are within the stated uncertainty range of standard materials. The expanded uncertainties were 0.2 mg·kg-1, 0.04 mg·kg-1, and 0.08 mg·kg-1 (K = 2) for magnesium, zinc, and copper, respectively. Therefore, this method has high trueness, good reproducibility, and simple operation and is suitable for tracing the values of magnesium, zinc, and copper in human serum.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.