Abstract

Blood flow in the superficial liver tissue was measured by recording the clearance rate of the β emissions of 85Kr following its injection into the portal vein. The flow values calculated from monoexponential clearance curves were compared to portal venous blood flow simultaneously recorded with a calibrated electromagnetic flow probe in six dogs whose hepatic artery had been ligated. The liver: blood partition coefficient of 85Kr was measured in these dogs and was found to be 0.91. There was a close correlation ( r = 0.95) between the blood flow values obtained from the two techniques over a wide range of flow values. This shows that measurement of liver blood flow in the surface layers of the liver using the 85Kr clearance method provides a true reflection of the blood flow in the liver tissue as a whole.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.