Abstract
Measurements of the performance of a miniature, portable 12-mm-diameter, 57-mm-length low-temperature cofired ceramic (LTCC) ion mobility spectrometer drift tube were undertaken to verify models of ion transport and determine the physical shape of the ion "swarms" in the LTCC tube. Simplified two-dimensional Gaussian models of ion swarm shape were fit to measured data to extract geometrical shape parameters. Results indicate that tube-transfer function effects that produce asymmetric ion swarms are minimized in the tube reducing temporal dispersion. Data are presented that illustrate the swarm shape as a function of gate time, electric field magnitude, and total charge in the ion swarm. Characterization and understanding of the ion transport mechanisms and effects that limit the resolution and other performance parameters of miniature IMS drift tubes is essential to the development of practical, robust, portable systems for "first responder" and homeland security missions.
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.