Abstract
Changes of corneal properties induced by laser in situ keratomileusis (LASIK) results in low inaccurate intraocular pressure (IOP) readings by Goldmann applanation tonometry (GAT). Before and after LASIK, the applied value of IOP, measured by dynamic contour tonometry (DCT) in comparison to GAT, was evaluated. Before and 1, 4 weeks after LASIK, the IOP in 30 cases (60 eyes) was measured by GAT and DCT respectively. The obtained results were statistically processed by SPSS11.5 statistical software. The results showed that central corneal thickness (CCT) could affect GAT measurements but not DCT measurements. The comparison of IOP one and 4 weeks after LASIK revealed that the readings from GAT was separately decreased by 5.00 +/- 1.12 and 5.45 +/- 1.13 mmHg as compared with those before LASIK, while those from DCT had no significant difference. It was concluded that LASIK-induced changes of CCT could influence the accuracy of GAT measurements, but had no influence on those from DCT. DCT was more beneficial to the measurements of IOP in normal eyes and those subject to LASIK surgery.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Journal of Huazhong University of Science and Technology [Medical Sciences]
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.