Abstract

Refraction contrast of simple objects obtained using diffraction enhanced imaging (DEI) was studied and compared to conventional radiographic contrast. Lucite cylinders and nylon wires were imaged using monochromatic synchrotron radiation at the National Synchrotron Light Source (nslsweb.nsls.bnl.gov/nsls/Default.htm) at the Brookhaven National Laboratory. The DEI images were obtained by placing a silicon analyser crystal tuned to the [333] diffraction plane in the beam path between the sample and the detector. To compare the DEI images with conventional radiographic images requires a consistent definition of refraction and absorption contrast. Conventional definitions of contrast favour conventional radiography and DEI contrast is defined to emphasize the specific characteristics of DEI. The proposed definitions were then used to find the DEI gain (the ratio of the DEI contrast with respect to the conventional image contrast). The results presented here show that the DEI gain is consistently greater than 1, indicating that DEI provides more contrast information than conventional radiography.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.