Abstract

Endoscopic retrograde cholangiopancreatography (ERCP) is widely used in the diagnosis and treatment of pancreatic and bile duct disorders. The procedure is performed under the guidance of fluoroscopy. This study aims to investigate the dose received by staff in the Endoscopy Department of AbuAli Sina Medicine and Organ Transplant Hospital in Shiraz. The dosimetry was performed using thermoluminescent dosemeters (TLD), type TLD-100. The values of Hp(10), Hp(3) and Hp(0.07) were estimated for the staff for 2months. According to the results obtained in this study, the equivalent dose of the gastroenterologist's body, eye lenses and hands was 0.045±0.001mSv, 0.111±0.014mSv and 0.357±0.034$\mathrm{mSv}$, respectively. This study showed that the annual radiation exposure for ERCP department staff of Abu Ali Sinai Hospital is less than the annual dose limit. However, if the principles of proper radiation protection and individual dosimetry are followed, the dose to staff members can be reduced.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.