Abstract
In this work, we propose a signal selection procedure for determination of hemoglobin (Hb) concentration in whole blood using near infrared (NIR) transmission spectral signals. A dataset of 190 whole blood NIR transmission spectra with reference Hb concentrations was used to evaluate the method. Spectral signals were selected based on the squared correlation coefficient (R2) between the signal and the Hb concentration. An improved uninformative variable elimination (UVE) procedure was performed to remove redundant signals from the primary selected signal set. A partial least squares (PLS) regression model was built with the final selected signals and the corresponding Hb concentrations. The results indicate that the proposed method is effective at increasing the predictive power of the NIR–PLS spectral model for determining Hb concentration in whole blood samples.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Biochemical and Biophysical Research Communications
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.