Abstract

The dimensioning of pressure relief devices (PRD) for cryogenic pressure equipment requires knowledge on the heat input at the maximum credible incident. In helium cryostats, this situation is typically defined by the loss of insulating vacuum (LIV), where the heat load is induced by desublimation and condensation of atmospheric air on the cryogenic surface. This surface is often covered with multi-layer insulation (MLI) in order to reduce the thermal radiation heat load in standard operation. During loss of insulating vacuum, the MLI represents a diffusive barrier for the air to reach the cryogenic surface, reducing the heat flux as well. Experimental reference data for the heat flux in case of LIV exist mainly for blank surfaces; only few data are published for MLI-covered helium surfaces. Therefore, the effect has been investigated in the cryogenic safety test facility PICARD at KIT. This paper presents the results of venting experiments carried out with different numbers of layers and different types of MLI.

Highlights

  • Experiments performed at PICARD with MLI Effect of multi-layer insulated helium cryostats was measured and evaluated Influence on manufacturing characteristics of the MLI demonstrated Relevant heat flux for dimensioning of PRD discussed Results differ from literature – higher heat flux for Type 1 MLI

  • Experiments performed at PICARD with MLI Effect of multi-layer insulated helium cryostats was measured and evaluated Influence on manufacturing characteristics of the MLI demonstrated (e.g. perforation holes) Relevant heat flux for dimensioning of PRD discussed Results differ from literature – higher heat flux for Type 1 MLI

Read more

Summary

Introduction

Results
Conclusion
Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call