Abstract

The exponential clearance rate constant, (kappa), and filtration fraction (FF) have been measured for dialysis of Gadolinium-diethylenetriaminepentaacetic acid (Gd-DTPA) (Magnevist, Berlex, Wayne, NJ) solutions by using a Look-Locker imaging technique under conditions of flow. The measured values of kappa for the Baxter CA-50, CA-110, and CA-210 filters were 0.0037 +/- 0.0003, 0.0057 +/- 0.0017, and 0.0092 +/- 0.0018 min-1, respectively, for dialysis of 4.0 liters of aqueous Gd-DTPA solutions. The measured values of FF for the Baxter CA-110 and CA-210 filters were 0.060 +/- 0.013 and 0.089 +/- 0.015, respectively, for dialysis of aqueous Gd-DTPA at 350 ml/min. The kappa and FF measurements agree with values that use inversion recovery (IR) on static samples obtained by drawing aliquots of solution during the course of dialysis. This in vitro experiment suggests that accurate in vivo measurements of filtration fraction and glomerular filtration rate (GFR) may be possible.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.