Abstract

A method for measuring the frequency stability of tunable laser is proposed by using confocal Fabry-Perot (F-P) interferometer. The F-P interferometer is used to get the output frequency of the laser as a reference, and the method eliminates the need of an independent optical source as a frequency reference. Using this technique, the frequency stability of the tunable external-cavity diode laser (ECDL) is measured to be 2.26×10−9 with an integration time of 20 ms.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.