Abstract
To assess the intraobserver and interobserver variability of transvaginal ultrasonographic measurement of endometrial thickness, two investigators (M.V., M.-F. D.) independently evaluated two successive measurements in 55 patients with a 5.0 MHz transvaginal transducer. Each observer was blinded to his or her own and to the other's results. Measurements were taken at the maximal thickness in the longitudinal uterine axis. The intraclass coefficients evaluating the agreement between two measures of a same observer were 94% and 97%; the coefficient was 94% between the two investigators. Intraobserver kappa values (agreement on endometrium < or = 5 mm and > 5 mm) were 0.70 and 0.81, respectively; the interobserver kappa value was 0.74. Transvaginal sonographic measurement of endometrial thickness has excellent intraobserver and good interobserver agreement. However, in clinical practice, an endometrial thickness measuring +/- 1 mm of a cut-off value should be interpreted with caution, especially when this may modify the clinical management.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Journal of ultrasound in medicine : official journal of the American Institute of Ultrasound in Medicine
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.