Abstract

PurposeMonitoring end-expiratory lung volume (EELV) is a valuable tool to optimize respiratory settings that could be of particular importance in mechanically ventilated pediatric patients. We evaluated the feasibility and precision of an intensive care unit (ICU) ventilator with an in-built nitrogen washout/washin technique in mechanically ventilated pediatric patients.MethodsDuplicate EELV measurements were performed in 30 patients between 5 kg and 43 kg after cardiac surgery (age, median + range: 26, 3–141 months). All measurements were taken during pressure-controlled ventilation at 0 cm H2O of positive end-expiratory pressure (PEEP).ResultsLinear regression between duplicate measurements was excellent (R2 = 0.99). Also, there was good agreement between duplicate measurements, bias ± SD: −0.3% (−1.5 mL) ± 5.9% (19.2 mL). Mean EELV ± SD was 19.6 ± 5.1 mL/kg at 0 cm H2O PEEP. EELV correlated with age (p < 0.001, r = 0.92, R2 = 0.78), body weight (p < 0.001, r = 0.91, R2 = 0.82) and height (p < 0.001, r = 0.94, R2 = 0.75).ConclusionThis ICU ventilator with an in-built nitrogen washout/washin EELV technique can measure EELV with precision, and can easily be used for mechanically ventilated pediatric patients.