Abstract

Assessing limits of agreement with helium dilution and repeatability of a new system (lung funcution, LUFU) that measures end-expiratory lung volume (EELV) in mechanically ventilated patients using the O(2) washin (EELV(Win)) and washout (EELV(Wout)) technique. LUFU consists of an Evita 4 ventilator, a side-stream oxygen analyzer, and a dedicated PC software. Prospective human study in a general ICU of a University hospital. Thirty-six mechanically ventilated patients. We obtained 36 couples of both EELV(Win) and EELV(Wout) measurements in each patient (5 with healthy lungs, 9 with ALI, 22 with ARDS). Measurements were obtained with patients ventilated either by assisted (ASB, 16 measurements) or controlled (CMV, 20 measurements) ventilation. In 19 of 20 cases in CMV, we obtained helium dilution measurements (EELV(He)). Bias for agreement with EELV(He) was -16 +/- 156 and 8 +/- 161 ml, respectively, for EELV(Win) and EELV(Wout). Bias for agreement between EELV(Win) and EELV(Wout) was 28 +/- 78 and 23 +/- 168 ml, respectively, for CMV and ASB. During CMV bias for repeatability were 8 +/- 92 and 23 +/- 165 ml, respectively, for EELV(Win) and EELV(Wout). During ASB bias for repeatability were 32 +/- 160 and -15 +/- 147 ml, respectively, for EELV(Win) and EELV(Wout). The LUFU method showed good agreement with helium, and good repeatability during partial and controlled mechanical ventilation. The technique is simple and safe.