Abstract

This cross-sectional study investigated the reproducibility of repeated elasticity and transepidermal water loss (TEWL) measurements with the DermaLab® on 32 active burn scars and healthy skin. Intra- and inter-observer reproducibility was examined by means of intra-class correlation coefficients (ICC) and standard error of measurements (SEM). Results showed good ICC values and rather high SEM values for inter- and intra-observer reproducibility of elasticity measurements. For TEWL measurements, ICC values were good and SEM values were high for inter- and intra-observer reproducibility. There was a significant difference between the estimated mean elasticity values of normal skin and grafted scars and between normal skin and spontaneously healed scars (p≤0.003). For the estimated mean TEWL values, there was a significant difference between normal skin and spontaneously healed scars (p=0.036). A significant negative relation was reported between mean TEWL and time after burn (p=0.008). In clinical trials it is necessary to interpret patient-specific changes in elasticity and TEWL with caution, since the SEMs of both modes are rather high. We therefore recommend the use of a mean of repeated measurements of elasticity and TEWL to decrease the SEM.